5 Habits to Avoid Cross-Contamination Between Two Measurement CampaignsField best practices

5 Habits to Avoid Cross-Contamination Between Two Measurement Campaigns

A sampling train that is poorly decontaminated between two sites can invalidate an entire campaign on parameters measured at trace levels. Here are the field habits that limit this risk.

Updated on
Share

An Underestimated Risk for Parameters Measured at Trace Levels

For pollutants such as dioxins, furans or heavy metals, measured at extreme trace levels, a residue left on a probe, filter holder or transfer line from a previous campaign can significantly skew the next result, well beyond the measurement uncertainty normally tolerated by the method. This cross-contamination risk is all the higher when several sites with different pollution profiles are visited back to back in the same week.

The Habits That Limit the Risk

Systematically dismantle the parts in contact with the sample (probe, filter holder, fittings) between two campaigns rather than relying on a quick rinse, particularly when the next site targets a different parameter from the previous one.

Go further

Standards mentioned

Frequently asked

FAQ

Questions about this topic? Here are the most common answers.

Another question?
Is a simple water rinse enough between two campaigns?

No, particularly for parameters measured at trace levels such as dioxins: a quick rinse does not guarantee the removal of residues adsorbed on surfaces. Complete dismantling and cleaning of the parts in contact with the sample remain necessary.

What is a field blank and why carry one out?

A field blank consists of running the sampling train under the real conditions of the campaign, without actually sampling any pollutant, to check for the absence of residual contamination before starting the final measurement.

A project to scope? Our experts can help.