First question: do you have an extraction network?

If your laboratory has an extraction network to the outside (or can install one), contaminated air can be discharged outside the building. Otherwise you need an air filtration and recirculation solution.

With extraction: fume cupboard or externally ducted hood

The [alarmed fume cupboard](/en/produits/laboratoire/sorbonne-laboratoire-alarme-rupture-confinement/) (NF XPX 15-206) permanently monitors air inlet velocity and triggers a visual and audible alarm in case of containment failure; its opening is limited to 400 mm above the worktop. The [externally ducted hood](/en/produits/laboratoire/hotte-laboratoire-raccordement-extraction-exterieure/) discharges air directly outside the building: air is not recycled in the room and there is no filter to monitor.

Without extraction: filtration hood or recirculating cupboard

The [filtration hood](/en/produits/laboratoire/hotte-filtration-laboratoire-charbon-actif-hepa/), compliant with NFX15-211, offers activated carbon filters to capture chemical vapours and HEPA filters to stop particles, with optional speed controller, alarm and filter saturation indicator. The [recirculating fume cupboard](/en/produits/laboratoire/sorbonne-recirculation-air-sans-conduit-exterieur/) filters the air drawn in at the workstation before returning it to the room, with no extraction duct.

What next?

The choice also depends on the products handled and their use. For the furniture around this equipment, see our guide [laboratory fit-out](/en/ressources/blog/amenager-laboratoire-paillasses-sorbonnes-hottes-armoires/) or [contact our experts](/en/contact/).

What we offer

Related instruments

Frequently asked

FAQ

Questions about this topic? Here are the most common answers.

Another question?
What is the difference between a fume cupboard and a filtration hood?

A fume cupboard (or externally ducted hood) discharges contaminated air outside through a duct; a filtration hood filters it (activated carbon, HEPA) then recycles it in the room.

Which standard applies to a filtration hood?

The CDL-Tecora filtration hood range complies with standard NFX15-211.

Does an externally ducted hood need filters?

No: air is discharged outside the building, with no internal filtration or saturation indicator to monitor.

A project to scope? Our experts can help.

Personalised advice and a fast quote for your measurement project.